Type K Calculator / Tolerance classes

Limits of error · IEC / ASTM

Type K Tolerance Classes

IEC 60584-1 and ASTM E230 limits of error, side by side, with an interactive pass/fail checker for as-found calibration data.

Thermometrics Corp →
Temperature at which the limit of error is evaluated.
Enter the measured error of the sensor under test to compare against the class band.
Permitted error band at this temperature
—

Request a quote with this spec →

Carries the class and temperature into the Thermometrics RFQ form.

How the classes are defined

Every class is the greater of a fixed °C floor or a fraction of the measured temperature: IEC Class 1 is ±1.5 °C or 0.4% (−40 to 1000 °C); Class 2 is ±2.5 °C or 0.75% (−40 to 1200 °C); Class 3 is ±2.5 °C or 1.5% (−200 to +40 °C, cryogenic). ASTM E230 Special is ±1.1 °C or 0.4% and Standard is ±2.2 °C or 0.75%, both 0 to 1260 °C.

The percentage term governs above the crossover temperature (floor ÷ fraction): 375 °C for IEC Class 1, 333 °C for Class 2, 275 °C for ASTM Special, 293 °C for ASTM Standard. Below the crossover the fixed floor applies. IEC Class 1 is broadly comparable to ASTM Special and Class 2 to Standard, but the floors, ranges, and crossovers all differ — a purchase order should name one standard explicitly rather than assume equivalence.

Type K permitted error band (±°C) by temperature and class
°CIEC Class 1IEC Class 2ASTM SpecialASTM Standard
0±1.50±2.50±1.10±2.20
100±1.50±2.50±1.10±2.20
200±1.50±2.50±1.10±2.20
300±1.50±2.50±1.20±2.25
400±1.60±3.00±1.60±3.00
500±2.00±3.75±2.00±3.75
600±2.40±4.50±2.40±4.50
700±2.80±5.25±2.80±5.25
800±3.20±6.00±3.20±6.00
900±3.60±6.75±3.60±6.75
1000±4.00±7.50±4.00±7.50
1100—±8.25±4.40±8.25
1200—±9.00±4.80±9.00
1260——±5.04±9.45

What the tolerance does and does not promise

A tolerance class bounds the initial error of new, essentially homogeneous wire. It says nothing about in-service behaviour: EMF aging, green rot, and shunting all accumulate on top of the class band.

Per the standard's own notes, the stated tolerances apply to new wire used within the recommended temperature limits, and the EMF–temperature relationship may change with time in use — results at manufacture do not necessarily hold over an extended service period. That is why loops with tight setpoints schedule periodic comparison against a reference standard; recalibration and certificates are covered under thermocouple calibration. Note also that the percentage terms are defined on temperatures expressed in °C, and the equivalent °F tolerance is 1.8× the °C value.

For QA work: enter the as-found deviation from your comparison check in the tool above and it returns a within/out-of-class verdict against the selected band. Out-of-class sensors in critical loops should be replaced rather than offset-corrected, since the drift mechanism that moved them rarely stops. Class 1 / Special-limits constructions are listed in the Type K catalog.